This work presents a non-invasive arterial blood pressure (ABP) waveform monitoring technique using ultrasound. A portable ultrasound system to excite ultrasound transducers and acquire data is designed with off-the-shelf components. The insonation angles are identified using a vector Doppler technique based on the cosine dependency of the Doppler signals. The pulse pressure of an estimated waveform at the left common carotid artery is compared to the standard sphygmomanometer measurement in a clinical test. The estimated carotid ABP waveform shows excellent agreement to the finger ABP waveform with expected discrepancy of the systolic peak shape due to different measurement sites. The proposed method also tracks slow blood pressure fluctuations. This validation on human subjects shows potential for a noninvasive blood pressure waveform monitoring device at central arterial sites.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.