Abstract

The purpose of the study was to compare cardiac output (CO) measurement by continuous (CTD) with that by conventional thermodilution (TD) in critically ill patients. In 19 of 20 critically ill patients requiring a pulmonary artery catheterism, 105 paired CO measurements were performed by both CTD and TD. Regression analysis showed that: CTD CO = 1.18 TD CO − 0.47. Correlation coefficient was 0.96. Bias and limit of agreement were — 0.8 and 2.4 L · min−1, respectively. When a Bland and Altman diagram was constructed according to cardiac index ranges, biases were −0.2 and −0.3 and −0.8 L · min−1 · m−2 and limits of agreement were 0.3, 0.7 and 1.6 L · min−1 · m−2 for low ( 4.5 L · min−1 · m−2) cardiac indexes, respectively. It is concluded that CTD, compared with TD, is a reliable method of measuring CO, especially when cardiac index is ≤4.5 L · min−1 · m−2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.