AbstractA molecularly imprinted polymer (MIP)‐based electrochemical sensor was developed for the monitoring of hydrochlorothiazide (HCT) in dialysate and hemodialysis wastewater. The MIP was constructed by the electropolymerization of ortho‐phenylenediamine on multi‐walled carbon nanotube modified glassy carbon electrode surface. The experimental optimization was carried out by Plackett‐Burman design screening and central composite design. Through square wave voltammetry (SWV) analysis, the sensor presented linear response in the range of 5.0×10−6 to 1.0×10−4 mol L−1 (R=0.9995) and detection limit of 3.2×10−6 mol L−1. The film morphology was investigated by SEM, EDX and AFM, and the electrochemical response was confirmed by EIS. The MIP film exhibited excellent selectivity for HCT, rapid response and good practicability. The sensor was successfully applied for direct and selective determination of HCT in dialysate and hemodialysis wastewater samples with recovery rates ranging from 97 % to 102 % based on the standard addition method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.