BackgroundArterial air embolism during open heart surgery may cause postoperative complications including cerebral injury, myocardial dysfunction, and dysrhythmias. Despite standard de-airing techniques during surgery large amounts of arterial air emboli may still occur, especially during weaning from cardiopulmonary bypass. To prevent this insufflation of carbon dioxide in the wound cavity has been used since the 1950s. The aim of this study was to assess a new mini-diffuser for efficient carbon dioxide de-airing of a minimal invasive cardiothoracic wound cavity model. Up until now no device has been evaluated for this purpose.MethodsA new insufflation device, a mini-diffuser, was tested. A thin plastic tube was used as control. The end of the mini-diffuser or the control, respectively, was positioned in a minimal invasive thoracic wound model. Remaining air content was measured during steady state and during intermittent suction with a rough suction device at different carbon dioxide flow rates. Measurements were also carried out in the open surgical wound during minimal invasive aortic surgery in six patients.ResultsThe air content was below 1% 4 cm below the surface of the open wound model during continuous carbon dioxide inflow of 2–10 L/min with the mini diffuser. In comparison, carbon dioxide insufflation via the open-ended tube resulted in a mean air content between 10 and 75%. The mean air content of the wound model remained below 1% at a carbon dioxide flow rate of 3–5 L/min during intermittent application of a suction device with a suction rate of 15 L/min. In 6 patients undergoing minimal invasive aortic valve replacement air content in the open surgical wound remained below 1% at a continuous carbon dioxide flow rate of 5 and 8 L/min via the mini-diffuser, respectively.ConclusionsThe mini diffuser was effective for carbon dioxide de-airing, i.e. < 1% remaining air, of a minimal invasive cardiothoracic wound cavity model with and without intermittent rough suction as well as in patients undergoing minimal invasive aortic valve surgery.
Highlights
Arterial air embolism during open heart surgery may cause postoperative complications including cerebral injury, myocardial dysfunction, and dysrhythmias
A very small but significant rise in air content occurred after 5 s for all flows (p < 0.001), after which the air content reversed to steady state levels after 10 s for all flows
Despite a two second intermittent rough suction force of 10 liters per minute (L/min) the air content remained below 1% at all Carbon dioxide (CO2) flow rates, except when the flow was only 2 L/min, which resulted in an increase of the air content to approximately 2.4% (p < 0.001)
Summary
Arterial air embolism during open heart surgery may cause postoperative complications including cerebral injury, myocardial dysfunction, and dysrhythmias. Despite standard de-airing techniques during surgery large amounts of arterial air emboli may still occur, especially during weaning from cardiopulmonary bypass. To prevent this insufflation of carbon dioxide in the wound cavity has been used since the 1950s. The aim of this study was to assess a new mini-diffuser for efficient carbon dioxide de-airing of a minimal invasive cardiothoracic wound cavity model. Up until now no device has been evaluated for this purpose. The aim of this study was to assess a new device, a mini-diffuser, for efficient CO2 de-airing of a minimal invasive cardiothoracic wound cavity model and in patients undergoing minimal invasive aortic valve surgery
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.