Abstract

Capnography is a noninvasive method of monitoring exhaled CO2and is used in many situations as a reflection of arterial CO2concentration. Expired CO2can be accurately measured using infrared spectrometry, mass spectrometry, and Raman scattering. Colorimetric CO2detectors are useful as a qualitive indicator of the presence of CO2. Trending end-tidal CO2and analysis of the capnographic wave form can provide information on changes in a patient's cardiopulmonary status or malfunction of equipment used to support a patient's cardiovascular or pulmonary systems. The accuracy of end-tidal CO2measurements as a reflection of arterial CO2tension in critically ill patients with respiratory or cardiovascular problems is limited. Changes in deadspace, respiratory rate and depth, positive end-expiratory pressure, areolar circulation, and cardiac output can significantly impact end-tidal CO2values more than arterial CO2concentrations. With marked changes in end-tidal CO2that are of concern, arterial blood gas analysis should be obtained to determine if the end-tidal CO2measurement is representatire of a change in arterial CO2concentrations. Despite the limitations of end-tidal CO2as a reflection of arterial CO2, it is still useful as a noninvasive monitor to alert clinicians of potential changes in a patient's cardiopulmonary condition.