Abstract
The performance of pyromellitic electrolyte for capillary zone electrophoresis of inorganic selenium species in the presence of selected common anions with indirect UV detection was investigated. The separation was achieved with pyromellitic electrolyte at pH 8.8 and hexamethonium hydroxide as the electroosmotic flow modifier. Obtained detection limits of 0.17 μg ml −1 for Se(VI) and 0.29 μg ml −1 for Se(IV) were improved by a factor of 5–7 in comparison with chromate electrolyte, which has been mainly employed for selenium analysis. Good resolution for nitrate–Se(VI) peaks were obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.