Abstract

Synthetic peptide samples may contain counter-ions such as acetate or trifluoroacetate as a result of their method of preparation. Furthermore, because acetic acid (HOAc) and trifluoroacetic acid (TFA) are frequently used reagents in peptide synthesis, these acids may be found in synthetic peptide samples as impurities. This paper describes a method validation to determine HOAc and TFA in synthetic peptide samples by capillary electrophoresis (CE) using an internal standard (I.S.) with indirect UV detection. Typical analytical parameters such as precision, linearity, accuracy, specificity, limit of detection and ruggedness were evaluated during the validation. In addition, the contents of HOAc and TFA in two synthetic opioid peptide samples, TIPP[ψ] and Orphanin FQ, were determined using the validated method. A unique feature of the method is that it offers determination of both acids in a single assay using a common I.S. The method is very efficient because of relatively short electrophoretic migration times (typically 2 to 8 min) for the acids investigated. This paper also discusses the factors that affect precision in a CE assay.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call