Abstract

To determine the frequency and causes for limitations in the radiographic evaluation of surgically treated long bone fractures. Six readers separately scored 140 sets of antero-posterior (AP) and lateral radiographs of surgically treated long bone fractures, using a radiographic union score (RUS). We determined the rate of assessability of the fracture edges at each of the four cortical segments (n=560) seen tangentially on the two radiographs and the causes for non-assessability. The rate of feasibility of the RUS (more than two fracture edges assessable per fracture) was determined and compared according to different parameters. Fracture edges were visible in 71% to 81% of the 560 cortical segments. Metal hardware superimposition was the most frequent cause for non-assessability (79-95%). RUS values could be calculated in 58% to 75% of fractures. Scoring was statistically significantly less frequently calculable in plated (31-56%) than in nailed fractures (90-97%), in distal (47-61%) than in proximal (78-89%) bones and in upper (27-49%) than in lower (76-91%) limb bones (P≤0.01). The type of stabilization hardware is the main limiting factor in the radiographic assessment of surgically treated long bone fractures. Scoring was feasible in only 31% to 56% of plated fractures.