Abstract

This paper describes a new alternative to conventional radiography system currently used for radiostereometric analysis studies. Instead of using two non-calibrated X-ray sources with a cumbersome calibration cage, we propose to use the biplanar radiography EOS system. Its fixed configuration provides a preliminary calibration and a much simpler acquisition protocol. A flexible and accurate calibration method is presented to optimize EOS default calibration using a simple object and a self-calibration method. To validate our system, we calculate the 3D reconstruction error of a known object. Results showed an accuracy of 70±11μm and 0.05±0.02° for translation and rotation respectively, and an average epipolar error of 23±03μm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.