Abstract

This paper presents the most recent mathematical modelling and procedure for Laser-Doppler Anemometers (LDA) calibration used by the Brazilian National Institute of Metrology Institute, the National Institute of Metrology, Quality and Technology (Inmetro). Here we calculate the fringe spacing calibration coefficient (C?) and the Burst Spectrum Analyser (BSA) calibration coefficient (C???). Using these two calibration coefficients we can calibrate a LDA in a determinate velocity range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.