Abstract
The purpose of this work was to evaluate differences in air-kerma rate of the older, S7500 water-cooled Xoft Axxent source and newer, S7600 Galden-cooled source. The Attix Free Air Chamber (FAC) at the UWMRRC was used to measure the air-kerma rate at 50 cm for six S7600 Xoft Axxent sources. The average measured air-kerma of the S7600 sources was compared with the measured average air-kerma rate from five S7500 sources. The air-kerma rates of the S7500 sources were measured in a Standard Imaging HDR 1000+ well chamber. The FAC measurements were used to determine a well chamber calibration coefficient for the S7600 source. The S7500 calibration coefficients were incorrectly applied to the S7600 sources to indicate the magnitude of error that can occur if the incorrect calibration coefficient is used. A 10.3% difference was observed between the average air-kerma rates of the two sources although a 17% difference was observed between their calibration coefficients. The application of the S7500 calibration coefficient to the S7600 sources resulted in measured air-kerma rates that were 20% greater than the true value. This work indicates the need for a new air-kerma rate standard for the S7600 sources, and the results presented in this work are indicative of values that would be obtained at National Institute of Standards and Technology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.