Abstract

A procedure for calibration of fluorescence signals from adult rat heart cells loaded with the -AM ester of fura-2 is described. Calibration is complicated by dye compartmentation and potentially incomplete dye hydrolysis. These problems were overcome by subtracting from fluorescence transients the non-cytosolic (mitochondrial) component of fura-2 fluorescence plus any Ca-insensitive component of dye fluorescence, after selectively and sequentially quenching cytosolic and non-cytosolic dye with Mn. The Kd of fura-2 in cells loaded by the -AM ester, in cells depleted of ATP and equilibrated with Ca buffers, was found to be 371 +/− 39 nM at 37°C. We found that calibration values for R MAX and R MIN derived from previously measured cells were of general validity, removing the need to measure R MAX and R MIN on every cell. Once these calibration values are determined, the calibration procedure to measure cytosolic Ca on any cell is a five minute procedure to determine compartmentation, using just one non-toxic and inexpensive solution. Finally, we have calculated how the errors intrinsic to the measurements translate into errors of the calculated Ca concentration and transient peak heights. These calculations allow reasonable parameters for data acquisition to be set.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.