Abstract

A method is proposed for measuring the concentration dependence of the thin-layer gas sensor sensitivity in a dynamic mode using only one certified gas-air sample with an impurity concentration corresponding to the upper limit of the concentration range for which the calibration is performed. A decrease in the total time of the measurement run is an advantage of the method. The response speed of the sensor for ethanol vapor in air on the basis of a thin tin dioxide (SnO2) layer is investigated, and the concentration dependence of its sensitivity to ethanol vapor is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.