Abstract

Confocal micro-beam X-ray fluorescence analysis (confocal micro-XRF) is a non-destructive analytical tool for investigation of sample composition that enables acquiring three-dimensionally resolved information. This work describes a calibration procedure of a laboratory confocal micro-XRF setup, which leads to determination of its characteristic parameters. The calibration is performed using a tabletop confocal micro-XRF spectrometer designed recently at the Czech Technical University in Prague. The calibration procedure performed within this work comprises the essential steps of the setup characterization: excitation spectrum calculation, experimental determination of energy-dependent confocal volume size and integral sensitivity and calculation of the spectrometer sensitivity function. The results of the setup calibration will be used for development of a procedure enabling quantitative evaluation of the measured depth profiles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.