Abstract

Abstract One of the challenges of measuring neutron fluences around medical linacs is the fact that the scattered photon fluence is important and higher than the surrounding neutron leakage fluence. Additionally most electron accelerators are pulsed, with repetition rates of the order of hundreds of Hertz, while the pulse duration is in the microsecond range. For this reason, neutron fluence around RT linacs is usually measured through passive methods, with the inconvenience of their time consuming analysis. A new neutron detector, based on the relation between Single Event Upsets (SEU) in digital SRAM memories and the existing thermal neutron fluence, has been developed. This work reports the calibration results of prototypes of this detector, obtained from exposures to the Physikalisch-Technische Bundesanstalt in Braunschweig (PTB) moderated 252 Cf source, to PTB quasi-monoenergetic neutron beams of 0.565 MeV, 1.2 MeV, 5 MeV, 8 MeV and 14.8 MeV, and to the GKSS thermal neutron beam.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.