Abstract

This paper describes the calibration of a 10 nF capacitance standard traced to the DC quantum Hall resistance primary standard using a multi-purpose digital impedance bridge. The bridge can directly compare the unlike impedances in the fully digital (FD) mode, so the capacitance standard can be calibrated by a direct comparison against the AC resistance standard. The measurement ability of the bridge in the FD mode was evaluated in various angles, and the bridge voltage dependencies were improved by selecting appropriate impedances for the high current arms and also by using an additional injection circuit. Thereby, the measurement accuracy could reach up to 10−7 in relative terms. Based on the bridge evaluation, a 10 nF capacitance standard was calibrated against a 10 kΩ resistance standard, and the measurement value agreed with a reference value in the level of 10−7.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.