Abstract
In recent decades, the calibration-independent methods have successfully been employed to determine the complex permittivity of dielectric materials without the need of calibration before measurements. However, these methods experience their respective problems from the requirement of two samples, the movement of the sample, or the introduction of extra assistant transmission lines. To eliminate these problems, a simple method is proposed based on transmission/reflection method in this paper. This method can afford high accurate measurement results without singularities by only two measurements of uncalibrated scattering parameters with one sample and one line. One measurement is for the empty fixture and the other for the same fixture holding the sample at a single position. Additionally, the principle of the proposed method ensures it independent of the fixture and the sample position. Some experiments were made using the vector network analyzer 8753ES and a Φ 16-mm coaxial line fixture. The proposed method is validated by the measurements of the permittivity of polytetrafluoro ethylene from 2 to 6 GHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Electromagnetic Compatibility
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.