Abstract

Cylindrical specimens may be tested advantageously by using grazing-incidence interferometry. A multiple positions test in combination with rotational averaging has recently been used to separate the surface deviations of the specimen from the interferometric aberrations. To reduce the measuring time and to check whether the results are reliable, a second procedure is now investigated, which uses the principle of the multiple positions test to determine quantities proportional to the difference quotients of the surface deviations. After numerical integration, the results can be compared with those obtained previously by rotational averaging. The measurement principle is described, and calibration results are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.