Abstract

Interferometry in grazing incidence can be used to test cylindrical mantle surfaces. The absolute accuracy of the resulting surface profiles is limited by systematic wavefront aberrations caused in the interferometer, in particular due to an inversion of the test wavefront in an interferometer using diffractive beam splitters. For cylindrical specimens, a calibration method using four positions has therefore been investigated. This test is combined with another method of optical metrology: the rotational averaging procedure. The implementation for grazing incidence is described and measurement results for hollow cylinders are presented. The gain in accuracy is demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.