Abstract

An ultraviolet (UV) optical absorption system has been developed for absolute calibrations of nitric acid (HNO3) and ammonia (NH3) permeation tube emission rates. Using this technique, dilute mixtures containing NH3 or HNO3, both of which interact strongly with many surfaces, are accurately measured at levels below a part per million by volume. This compact and portable instrument operates continuously and autonomously to rapidly (<1 h) quantify the emission of trace gases from permeation devices that are commonly used to calibrate air-monitoring instruments. The output from several HNO3 and NH3 permeation tubes, with emission rates that ranged between 13 and 150 ng/min, was examined as a function of temperature, pressure, and carrier gas flow. Absorptions of 0.015% can be detected which allows a precision (3sigma) of +/-1 ng/min for the HNO3 and NH3 permeation tubes studied here. The accuracy of the measurements, which relies on published UV absorption cross sections, is estimated to be +/-10%. Measurements of permeation tube emission rates using ion chromatography analysis are made to further assess measurement accuracy. The output from the HNO3 and NH3 permeation tubes examined here was stable over the study period, which ranged between 3 months and 1 year for each permeation tube.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.