Abstract
Absolute full-energy peak efficiencies, i.e. the ratio of the number of counts under the full-energy peak (FEP) to the number of photons at the same energy emitted by the source, are calculated for three true coaxial Ge(Li) detectors of differing size in the photon energy region 0.1–3.0 MeV. The detector volumes are 28.9, 82.1 and 144.9 cm 3. Three different source-to-detector geometries are considered, namely a point source, a needle-shaped source and an extended volume source. The attenuation effects in the sources, the aluminium windows of the detectors and the p-cores of the detectors are taken into account. Three different media are considered in the extended sources.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal Of Applied Radiation And Isotopes
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.