Abstract

An Ultrathin (~10 nm) InN ion sensitive field effect transistor (ISFET) with gate regions functionalized with phosphotyrosine (p-Tyr) is proposed to detect calcium ions (Ca 2+ ) in aqueous solution. The ISFET was miniaturized to a chip size of 1.1 mm by 1.5 mm and integrated at the tip of a hypodermic injection needle (18 G) for real-time and continuous monitoring. The sensor shows a current variation ratio of 1.11% with per decade change of Ca 2+ and a detection limit of 10-6 M. The response time of 5 sec. reveals its great potential for accomplishing fast detection in chemical and physiological sensing applications. The sensor would be applied in medical diagnosis and used to monitor continuous and real-time variations of Ca 2+ levels in human blood in the near future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.