Abstract

The computer-supported development of valves for cardiac-assist devices or artificial hearts is shown in relation to plastic technology. A CAD-system is used for the design development, whereas the dimensioning of the critical and highly stressed membranes is examined by FEM-analyses. Economic manufacture is permitted by the combined thermoforming-dip moulding technique; the blood-side components are made from biocompatible polyurethane to minimize blood damage. The first long-term results in the test set-up are compared to the FEM results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.