Abstract

Two-dimensional spatial localization can be achieved using a single nuclear magnetic resonance (NMR) pulse if the pulse is applied in the presence of a magnetic field gradient which is reorienting through two dimensions. The application of these pulses to whole-body NMR imaging systems is hampered, however, by amplitude and slew-rate constraints on the magnetic field gradients, resulting in longer pulses with relatively limited bandwidth. We describe here a variable rate method for shortening these pulses by as much as 36% for pulses limited by slew rate and 55% for pulses limited by gradient amplitude, without changing their spatial excitation profiles on resonance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.