Abstract
Blood samples of two cases were analyzed preliminarily by a classical spectrophotometric method (VIS) and by an automated headspace gas chromatographic method with nitrogen-phosphorus detector (HS-GC/NPD). In the former, hydrogen cyanide (HCN) was quantitatively determined by measuring the absorbance of chromophores forming as a result of interaction with chloramine T. In the automated HS-GC/NPD method, blood was placed in a headspace vial, internal standard (acetonitrile) and acetic acid were then added. This resulted in cyanide being liberated as HCN. The spectrophotometric (VIS) and HS-GC/NPD methods were validated on postmortem blood samples fortified with potassium cyanide in the ranges 0.5-10 and 0.05-5 mug/mL, respectively. Detection limits were 0.2 mug/mL for VIS and 0.05 mug/mL for HS-GC/NPD. This work shows that results obtained by means of the two procedures were insignificantly different and that they compared favorably. They are suitable for rapid diagnosis of cyanide in postmortem cases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.