Abstract
Wrinkles and air bubble artifacts may occur when preparing slides of semithin sections (0.5 μm) from blocks embedded in different resins. More than aesthetically annoying, wrinkles and air bubble artifacts may prohibit study of small structures. Present observations suggest that organic solvent based mounting media may interact with the resin of the section. This sometimes causes wrinkles and air bubble artifacts in the sections that degrade the quality of light microscope images. We compared the quality of semithin sections of several tissues in different resins using various types of mounting media. We observed that sections from Spurr's resin have many more artifacts. In particular, small, 2–10 μm round or oblong blister-like artifacts often plague our Spurr's resin sections. We demonstrate that Spurr's resin sections react with toluene and xylene in organic solvent based mounting media forming blisters, while sections from Araldite and L. R. White do not. We suggest combinations of embedding and mounting media for successful preparation of semithin sections for light microscopy without wrinkles, blisters, or air bubble artifacts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.