Abstract

A hydride generation procedure, via flow injection, coupled to electrothermal atomic absorption spectrometry was optimised for Bi determination in sea water and hot-spring water and acid extracts from coal, coal fly ash and slag samples. The effects of several variables such as hydrochloric acid and sodium tetrahydroborate concentrations, hydrochloric acid and sodium tetrahydroborate flow rates, reaction coil length, trapping and atomisation temperatures, trapping time and the Ar flow rate have been investigated by using a 2 9*3/128 Plackett–Burman design. From these studies, certain variables (sodium tetrahydroborate concentration and trapping time) showed up as significant, and they were optimised by a 2 2+star central composite design. In addition, a study of the bismuthine trapping and atomisation efficiency from graphite tubes (GTs) permanently treated with uranium, tantalum, lanthanum oxide, niobium, beryllium oxide, chromium oxide and tantalum carbide were investigated. The results obtained were compared with those achieved by iridium and zirconium-treated GTs. The best analytical performances, with characteristic mass of 35 pg and detection limit of 70 ng l −1, were achieved by using U-treated GTs. Accuracy were checked using several reference materials: 1643d (Trace Elements in Water), TM-24 (Reference Water), GBW-07401 (Soil) and 1632c (Trace Elements in Coal).