Abstract

A highly sensitive electrochemical biosensor was manufactured with triple synergistic catalysis to detect hydrogen peroxide (H2 O2 ). In this study, a highly sensitive biosensor based on Prussian blue-chitosan/graphene-hemin nanomaterial/platinum and palladium nanoparticles (PB-CS/HGNs/Pt&Pd biosensor) was fabricated for the detection of H2 O2 . The materials described above were modified on the electrode surface and applied to catalyze the breakdown of hydrogen peroxide. The current response of the biosensor presented a linear relationship with H2 O2 concentration from 6 × 10-2 to 20μM (R2 = 0.9766) and with the logarithm of H2 O2 concentration from 20 to 9×103 μM (R2 = 0.9782), the low detection limit of 25nM was obtained at the signal/noise (S/N) ratio of 3. Besides, the biosensor showed an outstanding anti-interference ability and acceptable reproducibility. PB-CS/HGNs/Pt&Pd electrodes are effective in measuring H2 O2 from living tumor cells, which implies that the biosensor has the potential to assess reactive oxygen species in various living tumor cells.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.