Abstract

The objective of this study was to evaluate the effects of a semi-constrained artificial disc with upward instantaneous center of rotation (ICR) on the biomechanics of the cervical spine. A three-dimensional nonlinear finite element model of the lower cervical spine (C4–C7) was developed using computed tomography (CT) data. The FE model was validated by comparing it to previously published experimental results for flexion-extension, lateral bending and axial rotation movements. The validated model was then altered to include prosthesis at the C5–C6 level. A hybrid test protocol was used to investigate the effects of total disc replacement. The results of this study showed that this artificial disc can help maintain the same range of motion (ROM) and intradiscal pressure as the intact model for most loading conditions. We also found that loads on the facet joints increased dramatically at index level. The capsular ligaments were also found to transmit more tension during flexion at implanted level. Although the artificial disc with upward ICR was found to restore normal kinematics, and prevented increases in intradiscal pressure, it was also associated with an overloading of the facet joints and capsular ligaments leading to potentially undesirable outcomes in the long term.