Study DesignThe study examined and compared four artificial cervical disks using validated finite element models.PurposeTo compare and contrast the biomechanical behavior of four artificial cervical disks by determining the external (range of motion) and internal (facet force and intradiscal pressure) responses following cervical disc arthroplasty (CDA) and to elucidate any device design effects on cervical biomechanics.Overview of LiteratureDespite CDA’s increasing popularity most studies compare the CDA procedure with anterior cervical discectomy and fusion. There is little comparative evaluation of different artificial disks and, therefore, little understanding of how varying disk designs may influence spinal biomechanics.MethodsA validated C2–T1 finite element model was subjected to flexion-extension. CDAs were simulated at the C5–C6 level with the Secure-C, Mobi-C, Prestige LP, and Prodisc C prosthetic disks. We used a hybrid loading protocol to apply sagittal moments. Normalized motions at the index and adjacent levels, and intradiscal pressures and facet column loads were also obtained.ResultsThe ranges of motion at the index level increased after CDA. The Mobi-C prosthesis demonstrated the highest amount of flexion, followed by the Secure-C, Prestige LP, and Prodisc C. The Secure-C demonstrated the highest amount of extension, followed by the Mobi-C, Prodisc C, and Prestige LP. The motion decreased at the rostral and caudal adjacent levels. Facet forces increased at the index level and decreased at the rostral and caudal adjacent levels following CDA. Intradiscal pressures decreased at the adjacent levels for the Mobi-C, Secure-C, and Prodisc C. Conversely, the use of the Prestige LP increased intradiscal pressure at both adjacent levels.ConclusionsWhile all artificial disks were useful in restoring the index level motion, the Secure-C and Mobi-C translating abilities allowed for lower intradiscal pressures at the adjacent segments and may be the driving mechanism for minimizing adjacent segment degenerative arthritic changes. The facet joint integrity should also be considered in the clinical decision-making process for CDA selection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.