Abstract

The pterional approach is the most popular surgical technique in aneurysm and skull base tumor removal. Reconstruction of the temporal contour deformity due to craniotomy requires graft implantation. Porous high-density polyethylene (PHDPE) as a craniofacial and pterional implant material recently became available. However, material properties of the pterional implant are not yet known. In order to measure the biomechanical properties of PHDPE, we implemented the tensile test, the three-point bending test and the water displacement method for density measurement. Elastic modulus varies from 227 to 307 MPa. Density range is 0.68 and 0.7 depending on the size of pores. The data can be used to study the character of the porous high-density polyethylene implant, how it resists stress or fatigue in combination with conventional plating systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.