Abstract
BackgroundPatients with a simple transversal fracture of the olecranon are often treated with a tension band wiring (TBW), because it is known as a biomechanically appropriate and cost-effective procedure. Nevertheless, the technique is in detail more challenging than thought, resulting in a considerable high rate of implant-related complications like k-wire loosening and soft tissue irritation. In the literature, a distinction is generally only made between transcortical (bi-) and intramedullary (mono-) fixation of the wires. There is the additional possibility to fix the proximal bent end of k-wire in the cortex of the bone and thus create a tricortical fixation. The present study investigates the effectiveness of bi- and tricortical k-wire fixation in a biomechanical approach.MethodsTBW of the olecranon was performed at 10 cadaver ulnas from six donors in a usual manner and divided into two groups: In group 1, the k-wire was inserted by bicortical fixation (BC), and in group 2, a tricortical fixation (TC) was chosen. Failure behavior and maximum pullout strength were assessed and evaluated by using a Zwick machine. The statistical evaluation was descriptive and with a paired t test for the evaluation of significances between the two techniques.ResultsThe average age of the used donors was 81.5 ± 11.5 (62–92) years. Three donors were female, and three were male. Ten k-wires were examined in BC group and 10 in the TC group. The mean bone density of the used proximal ulnas was on average 579 ± 186 (336–899) HU. The maximum pullout strength was 263 ± 106 (125–429) N in the BC group and increased significantly in the TC group to 325 ± 102 (144–466) N [p = .005].ConclusionThis study confirms for the first time biomechanical superiority of tricortical k-wire fixation in the olecranon when using a TBW and may justify the clinical use of this method.
Highlights
Patients with a simple transversal fracture of the olecranon are often treated with a tension band wiring (TBW), because it is known as a biomechanically appropriate and cost-effective procedure
The aim of the present study is the investigation of the biomechanical pullout strength of the k-wire fixation while using TBW to answer the question, whether a tricortical fixation is more stable than a bicortical fixation
Ten k-wires were examined in bicortical fixation (BC) and 10 in the tricortical fixation (TC) group
Summary
Patients with a simple transversal fracture of the olecranon are often treated with a tension band wiring (TBW), because it is known as a biomechanically appropriate and cost-effective procedure. The major disadvantages of these techniques are postoperative irritations caused by the implanted hardware and secondary dislocations of the k-wires [1, 4, 10, 14]. This can cause consecutive irritation or perforation of the skin and in the end a soft tissue infection [3]. Painful prominences of the k-wires are one of the most frequent postoperative complications and result in about 80% an implant removal
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have