In implant technology, open porous Ti coatings are applied as functional surface layers on prosthetic devices to improve osseointegration. Since a successful clinical performance strongly depends on the (initial) quality of bone ingrowth in the porous structure, surface functionalization of the porous Ti to incorporate an additional osteoconductive capacity is recommended. In this paper, a bioactive glass–ceramic coating is applied into the open porous network of Ti coatings with a pore throat size of 1–20μm through a sol–gel process. Using an all-alkoxide precursor route, homogeneous amorphous powders of three- (SiO2–CaO–P2O5) and four-component (SiO2–CaO–Na2O–P2O5) bioactive glass compositions are prepared. By sol impregnation followed by a heat treatment, it is possible to deposit a micrometer thin bioactive glass–ceramic layer on the walls of the internal pore surface, while the original porosity and the open pore structure of the Ti coatings are maintained. The tensile adhesion strength of the Ti/bioactive glass–ceramic composite coatings is 22 to 29MPa, suggesting a good mechanical adhesion.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.