Abstract
Main textThis report gives the result of a bilateral comparison of resistance between the NSAI-NML (Ireland) and the BIPM carried out from November 2020 to March 2021. Two 1 Ω and two 10 kΩ travelling standards belonging to the BIPM were used. The comparison was carried out with an 'A-B-A' pattern of measurements; the standards were measured first at the BIPM for a period of about one month, then for a period of about one month at the NSAI-NML, and finally again at the BIPM. The measurand was the 4 terminal dc resistance at low power. The BIPM was the pilot laboratory, and the comparison forms part of the ongoing BIPM key comparisons BIPM.EM-K13.a (for 1 Ω) and BIPM.EM-K13b (for 10 kΩ).The results from the NSAI-NML and the BIPM were found to be in good agreement, with a difference D smaller than the relative expanded uncertainty U c (95% confidence, k = 2) giving D = - 0.081 × 10-6, U c = 0.180 × 10-6 for 1 Ω and D = 0.069 × 10-6, U c = 0.424 × 10-6 for 10 kΩ.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.