Abstract

This report gives the result of a bilateral comparison of resistance between the NMC A*STAR (Singapore) and the BIPM carried out in 2018-2019. Two 1 Ω and two 10 kΩ travelling standards belonging to the BIPM were used. The comparison was carried out with an 'A-B-A' pattern of measurements; the standards were measured first at the BIPM for a period of about two months, then for a period of about one month at the NMC A*STAR, and finally again at the BIPM. The measurand was the 4 terminal dc resistance at low power. The BIPM was the pilot laboratory, and the comparison forms part of the ongoing BIPM key comparisons BIPM.EM-K13.a (for 1 Ω) and BIPM.EM-K13b (for 10 kΩ).The results from the NMC A*STAR and the BIPM were found to be in good agreement, with a difference smaller than the relative expanded uncertainty (95% confidence, k = 2) of 0.064 × 10−6 for 1 Ω and 0.036 × 10−6 for 10 kΩ.Main textTo reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.