Abstract

A strip-type glucose biosensor, prepared using screen-printing technology and comprising glucose oxidase (E.C.1.1.3.4.), peroxidase (E.C.1.1.3.13.) and ferrocyanide as mediator incorporated into graphite-hydroxyethyl cellulose matrices is described. The sensor acted at 0·0 V vs Ag/AgCl electrode, and the response time was 50–60 s. The calibration was linear up to 25 m m of glucose. The sensor response was constant in the range of pH 7·0–8·5. At 25°C the biosensor temperature coefficient was 2·7% °C −1. The sensor was insensitive to a physiological level of ascorbic acid (40 μ m and was used for glucose determination in whole blood.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.