Abstract

Biconvex posterior chamber lenses have optical advantages and decrease the risk of capsular opacification, but they are more likely to be pitted during Nd:YAG capsulotomy because of apposition of the lens to the capsule. This study reports the likelihood of surface damage to different formulations of poly(methyl methacrylate) at the energy levels required to open posterior capsules. Molded lenses are more easily damaged than higher molecular weight lathe-cut materials (P <.01), as expected. however, by keeping energy output low, even injection-molded lenses showed minimal damage, with mean pit size 39 ± 39 µm at 1 mJ. By using a converging contact lens, low power, and keeping the focus behind the capsule, damage to all materials tested should be clinically insignificant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.