Abstract

PurposeThis study aims to validate the dosimetric characteristics of High Dose Rate (HDR) 60Co source (Co0.A86 model) using GATE Geant4-based Monte Carlo code. According to the recommendation of the American Association of Physicists in Medicine (AAPM) task group report number 43, the dosimetric parameters of a new brachytherapy source should be verified either experimentally or by Monte Carlo calculation before clinical applications. The validated 60Co source in this study will be used for the simulation of intensity-modulated brachytherapy (IMBT) of vaginal cancer using the same GATE Geant4-based Monte Carlo code in the future. Materials and methodsGATE (version 9.0) simulation code was used to model and calculate the required TG-43U1 dosimetric data of the 60Co HDR source. DoseActors were defined for calculation of dose rate constant, radial dose function, and anisotropy function in a water phantom with an 80 cm radius. ResultsThe dose rate constant was obtained as 1.070±0.008cGy.h−1.U−1 which shows a relative difference of 2.01% compared to the consensus value, 1.092 ​cGy.h−1.U−1. The calculated results of anisotropy and radial dose functions starting from 0.1 cm to 10 cm around the source showed excellent agreement with the results of published studies. The mean variation of the radial dose and anisotropy functions values from the consensus data were 1% and 0.9% respectively. ConclusionFindings from this investigation revealed that the validation of the HDR 60Co source is feasible by the GATE Geant4-based Monte Carlo code. As a result, the GATE Monte Carlo code can be used for the verification of the brachytherapy treatment planning system.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call