Abstract

In spectroscopic ellipsometry (SE) measurements of thin films with inhomogeneous optical thickness, the spot size of the probe beam on the sample surface is a highly critical parameter in the treatment of the experimental data. SE measurements were performed on transparent samples prepared by various techniques: plasma enhanced chemical vapor deposition of diamond film, spin-coated PMMA films and thermally grown SiO 2 film. The first two deposition techniques can produce films with inhomogeneous optical thickness. By comparing SE measurements using different set-up configurations, we demonstrate that a combination of lenses and small diaphragms is most suitable for minimizing the effects related to the optical thickness distribution and light beam divergence. In this way mistakes are avoided in extracting physical parameters from SE optical modeling of these films.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.