Abstract

The beam diagnostic system of the TRISTAN Accumulation Ring consists of beam position monitors, visible radiation monitors, x-ray monitors, tune measurement setup, etc.. Eighty-six position monitors are installed around the ring. For the closed orbit measurement, a superheterodyne circuit is used to pick up the 479-th harmonic of the revolution frequency (795 kHz) out of beam pulse trains. Synchrotron light is observed in the visible region and in the x-ray region. Visible radiation is used in three ways: profile monitoring by TV cameras, beam current measurement and bunch shape observation by a streak camera. In the x-ray channel a multi-wire ionization chamber is used to get a digitized profile of the x-ray source. Stripline pickxps are installed to detect transverse oscillations of e+ and e- bunches. The envelope signal of pulse trains will be sent to an FFT processor for tune number identification. At the same time the signals are amplified and fed back to wideband deflection electrodes for damping of the oscillaitons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.