Abstract
This work reports on the results obtained for VIC-T dosimeter for radiotherapy dosimetry (17 % (w/v) N-vinylpyrrolidone, 8 % (w/v) N,N′-methylenebisacrylamide, 12 % (w/v) tert-butyl alcohol, 5 % (w/v) gelatine, 0.02 % (w/v) hydroquinone and 14 mM tetrakis (hydroxymethyl) phosphonium chloride (THPC)) measured with a spiral computed tomography (CT). Optimal spiral CT protocol, CT measurements conditions and image processing are provided. Basic features of VIC-T are discussed such as a calibration relation and dose resolution with respect to former N-vinylpyrrolidone containing dosimeters. Both adopted data processing performed with a new polyGeVero®-CT software package and its features are reported.