Abstract

Bacterial colony counts in reusable water reservoirs in a humidified oxygen supply system maintained only by the addition of nonsterilized water, with or without washing (conventional procedure), were examined. Bacteria were detected in all the reservoirs even before the operation of the oxygen suppliers, and colony counts of more than 1 x 10(2)/ml of bacteria were detected in some reservoirs without washing. Many bacteria were also detected in many reservoirs, after the operation of the oxygen suppliers, indicating that bacteria had contaminated the water reservoirs and propagated during this operation. These results suggest that it is not recommended to maintain these water reservoirs by the conventional procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.