Abstract

Temperature of X-ray tube is very high in operation. Therefore, information of detailed temperature distribution is indispensable for developing of the new X-ray tube. This study is conducted to obtain the detailed temperature distribution on the basis of the unsteady numerical analysis. Temperature near the target center became nearly 2000K immediately after the input. Maximum temperature with intensity distribution at the target was about 2310K. The maximum temperature was about 20% higher than uniform input case for target. Temperature gradient was steep near the target, and temperature was about 700K at a distance of 1 mm from the target center.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.