Abstract

The basic characteristics of the axial-symmetric transmission gauge (AT gauge), that were modified for measurement over a wide pressure range, are reported. A Faraday cup ion collector is used instead of a secondary electron multiplier to extend the measurable pressure range to 10−3Pa. The sensitivity of the gauge was determined by a conductance modulation method, and by direct comparison against an extractor gauge and a spinning rotor gauge as well. When operating in O2, there was systematic disagreement between pressure readings with the AT gauge and the extractor gauge. This is due to the difference in the elimination efficiency of the electron stimulated desorption ion.