Abstract

The axial length was measured in five hundred consecutive cases using the Kretz 7200 MA ultrasound unit and an immersion technique. The reproducibility of the measurements was within ±0.2 mm when performed manually, and within ±0.05 mm when performed electronically. Intraocular lens power calculations were performed using a modification of Colenbrander's formula. We predicted within +1 diopter in 78.8% of the cases. The accuracy was up to 83% when the axial length ranged between 23.0 and 25.0 mm. Stronger power lenses were implanted in shorter eyes and weaker power lenses in longer eyes, necessitating the use of a fudge factor. Surgeons using an immersion technique for axial length measurements should use formulas yielding stronger power lenses, such as Binkhorst's formula or our modification of Colenbrander's formula, with a fudge factor for short and long eyes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.