Abstract
To examine the relationship between axial length and keratometry measurement errors and intraocular lens (IOL) power calculations for pediatric eyes. The sensitivity of IOL power calculation to errors in axial length and keratometry measurements was computed as a function of axial length and keratometry for the SRK II, Hoffer Q, Holladay I, SRK/T, and Haigis formulas. The sensitivity of the IOL power calculation to an axial length measurement error is increased at 4 to 14 D/mm error in axial length in children compared with 3 to 4 D/mm error in axial length in adults. The error in calculation is 0.8 to 1.3 D/D error in keratometry measurement for both children and adults. Axial length measurement errors in pediatric eyes may lead to large errors in IOL power calculations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of American Association for Pediatric Ophthalmology and Strabismus
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.