Abstract
Prior to use in the ZEUS calorimeter, we measured the performance of 4800 photomultiplier tubes of the R580-12 type. The measurements were carried out under computer control using an automatic test system in which 80 tubes were tested simultaneously. The tests were made regarding the dark current, gain, linearity, long-term stability and gain stability under different intensities of dc background light. The results were used to both select and classify the tubes. Almost all of the tubes were found to be satisfactory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.