Abstract
A fully automated and highly sensitive method with a semi-microcolumn liquid chromatography system for the determination of rat plasma catecholamines (CAs) was developed. Automated on-line extraction of CAs in diluted plasma using a precolumn packed with strong acidic cation exchange resin was coupled with separation of CAs on a semi-microcolumn (250 x 1.5 mm id). fluorogenic derivatization with ethylenediamine and finally postcolumn peroxyoxalate chemiluminescence detection utilizing bis[2-(3,6,9-trioxadecanyloxycarbonyl)-4-nitrophenyl]oxalate (TDPO) and hydrogen peroxide. The detection limits were 0.91, 0.36 and 1.1 fmol for norepinephrine (noradrenaline), epinephrine (adrenaline) and dopamine, respectively, at a signal-to-noise ratio of 3. A good linearity of the calibration curve for each CA was observed in the range of 5.0 to 500 fmol for each CA using N-methyldopamine (N-MeDA) as an internal standard. The RSD for the proposed method (n = 5) were 3.7-9.5% for the intra-day assay and 6.6-10.0% for the inter-day assay. The volume of rat plasma required for the determination of CAs was 10 microliters.