Abstract

An automatic system was developed to prepare calibrating solutions for elemental measurements by inductively coupled plasma mass spectrometry (ICP-MS). The flow-rates of diluting solution, internal standard solution, and initial elementals solution were monitored with a single ultrasonic flow sensor. Multiple calibrating solutions prepared by the present automatic system were analyzed by ICP-MS, whose results provided a correlation coefficient better than 0.999 for calibration curves. The results for multi-elements in a river-water certified reference material (NMIJ CRM 7202-b) agreed with the certified values.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call