Abstract

An automatic flow-injection (FI) system including on-line separation by microdistillation and spectrophotometric detection has been developed for the determination of trace amounts of fluoride. This ion was separated from sample matrix by distillation in the presence of sulfuric and phosphoric acids, and was subsequently determined with spectrophotometry based on the mixed-ligand complex of lanthanum(III)–fluoride–alizarin complexone. The proposed FI system has high sampling frequency (20 samples h −1), small sample size (600 μl) and the dynamic range of 0.05–15 mg l −1 with relative standard deviations of below 1.2%. Interfering ions such as aluminum(III) and iron(III) was effectively eliminated. The method was successfully applied to the determination of fluoride in industrial drainage after water treatment.